Project 3: Radiofrequency (RF) Pulse Development and Design for High Field MRI
Project Leader: Gerald B. Matson, Ph.D.; UCSF and VA Medical Center, San Francisco

The promise of improved MRI results at high field strength is compromised by the difficulties encountered at high field, including: i) Non-uniform excitation, due to the non-uniform B1 field inherent at high field. Typically, the non-uniform excitation produces non-uniform tissue contrast, although other deleterious effects can be produced as well. ii) Large susceptibility gradients, which can distort slice positions unless large slice-select gradients are used. However, the limited RF power available on high field systems severely limits the gradient strength that can be used for T2-weighted images. The specific aims propose the further development and refinement of two new RF pulse designs to ameliorate these deleterious effects. In addition, further development of software for simulating MRI experiments is proposed to aid in effective implementation of these new RF pulses into suitably re-designed MRI experiments.